Condition: Refurbished, Certified Pre-owned
Wavelength range: 189-900 nm
Grating area: 36 x 185 mm
Diffraction grating: 79 lines/nm, blaze angle 76°, automated wavelength and slitsize control
Reciprocal linear dispersion: 2.4 nm/mm
Focal length: 300 mm
Spectral bandpass: 0.15 (at 200 nm)
Fused-quartz prism: 95 x 45 mm, 60°
Dimensions: 70 x 65 (0.46 m 2) x 65 cm (W x D x H)
Weight: 49 kg
Power: 100-230 V (±10%), 50/60 Hz (±1%), 300 VA (maximum)
Technical: Classified as a laboratory instrument. Complies with the applicable European Union directives and standards for safety and electro-magnetic compatibility for CE Marking, the safety requirements for Canada and the United States for CSA/NRTL certification and the FCC requirements for radio-frequency emissions. The instrument was developed and produced in compliance with ISO 9001.
Environmental: Dust-free, free of vibrations, ambient temperatures: +15 ̊C to +35 ̊C with a change rate of a maximum 3 ̊C per hour. Relative humidity: 20% to 80% non-condensing.
Optional: Optical cell (p/n 85576588), Perkin Elmer AS 93 Plus Autosampler (p/n B3140330), Quantachrome Quartz Rod (p/n 01536-6345), Perkin Elmer oil extractor (p/n N0775325), Cell- Nova- Long 9mm Large Bulb (p/n 74064), NE cell 9mm large NV/quadras (p/n 02137-7736), 9mm pellet long cell (p/n 74099), long pass filter-305 red (p/n 83228700), sample cell funnel black delrin (p/n 04000-3595), Perkin Elmer Lumina lamps: FE (p/n N305-0126), La (p/n N305-0141), K (p/n N305-0204), Ca-Mg-Zn (p/n N305-0208), W (p/n N305-0187), Co (p/n N305-0118), Li (p/n N305-0142) & PE EDL System 2 Lamp Element RB (p/n N305-0664).