Unlock Exclusive Discounts & Flash Sales! Click Here to Join the Deals on Every Wednesday!

Agilent 1100 G1377A Micro-WPS Well Plate Sampler, Agilent Technologies (CAT#: STEM-CPO-0894-ZJF)

Highlights

Refurbished to meet manufacturer's specification
In excellent condition
Thoroughly cleaned

Cat Number: STEM-CPO-0894-ZJF

Application: For High Performance Liquid Chromatography (HPLC)

Model: G1377A

Add to Cart



Description

The manufacturer of the Agilent 1100 G1377A Micro-WPS Well Plate Sampler is Agilent Technologies.
The Agilent 1100 G1377A Micro-WPS is a micro-well plate sampler for use with the Agilent 1100 Series HPLC system. It allows for automated liquid handling and sample injection into the HPLC system, and is designed for use with 96- or 384-well microplates. The Micro-WPS features a compact design and easy operation, making it ideal for use in high-throughput laboratory settings.

Specification

Condition: Pre-owned
Glp features: Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications: Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection range:
8 µl in 0.01 µl increments with the small loop capillary
40 µl in 0.01 µl increments with the extended loop capillary
Precision:
Typically < 0.5 % RSD of peak areas from 5 – 40 µl,
Typically < 1 % RSD from 1 – 5 µl
Typically < 3 % RSD from 0.2 – 1 µl
Sample viscosity range: 2 – 5 cp
Sample capacity:
2× well-plates (MTP) + 10 × 2 ml vials
100 x 2 ml in one tray
40 x 2 ml in half tray
Injection cycle time:
Typically, < 30 s using following standard conditions:
Default draw speed: 4 µl/min
Default eject speed: 10 µl/min
Injection volume: 0.1 µl
Carry-over:
Typically < 0.05 % using the following conditions:
Column: 150 x 0.5 mm Hypersil ODS, 3 µm
Mobile phase: Water/Acetonitrile = 85/15
Column Flow rate: 13 µl/min
Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test carryover
Outside wash of needle before injection: 20 sec with water using flush port

Features

Refurbished to meet manufacturer's specification
In excellent condition
Thoroughly cleaned
Certified with warranty

Related Products

Agilent 1100 G1364D Micro-FC Fraction Collector, Agilent Technologies
Agilent 1100 Series Isocratic Pump (G1310A), Agilent Technologies
Agilent 1100 Series Quaternary Pump (G1311A), Agilent Technologies
Advertisement