Unlock Exclusive Discounts & Flash Sales! Click Here to Join the Deals on Every Wednesday!

Agilent 1100 Series - G1361A HPLC Preparative Pump, Agilent Technologies (CAT#: STEM-C-0046-LGZ)

Highlights

1. Up to 100 mL/min flow is provided at 400 bar without changing the pump head.
2. For preparation and analytical applications at low flow rates (0.5 mL/min, upward), providing an accurate high-pressure binary gradient structure with low delay volume (700 µL).
3. Extended seal life due to the use of peristaltic pump automatic cleaning seal.

Cat Number: STEM-C-0046-LGZ

Application: For preparative and analytical applications with low flow rates.

Model: 1100 Series - G1361A

Add to Cart



Description

The manufacturer of Agilent 1100 Series - G1361A HPLC Preparative Pump is Agilent Technologies. Model is 1100 Series - G1361A. For preparation and analytical applications at low flow rates (0.5 mL/min, upward), providing an accurate high-pressure binary gradient structure with low delay volume (700 µL)。

Specification

Condition: used
Hydraulic system: Dual pistons in parallel
Settable flow range: V0.001 – 100 ml/min
Flow precision: < 0.5 % RSD
Pressure range: 20 to 400 bar (5880 psi) system pressure
Compressibility compensation: User-selectable, based on mobile phase compressibility
Recommended pH range: 1.0 – 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel.
Control and data evaluation: Agilent ChemStation for LC
Communications: Controller-area network (CAN), RS-232, APG
Remote: ready, start, stop and shut-down signals, CAN-DC OUT, LAN optional
Safety and maintenance: Extensive diagnostics, error detection and display (through
control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
GLP features: Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with user-settable limits and feedback messages. Electronic records of maintenance and errors.
Housing: All materials recyclable
Weight: 15.0 kg
Dimensions
(Height x Width x Depth): 200 x 345 x 440 mm
(8 x 13.5 x 18 inches)
Line voltage: 100 – 240 VAC, ± 10 %
Line frequency: 50 – 60 Hz, ± 5 %
Power consumption: 250 VA
Ambient operating temperature: 4 – 40 °C (41 – 104 °F)
Ambient non-operating
temperature: -40 – 70 °C (-4 – 158 °F)
Humidity: < 95 %, at 25 – 40 °C (77 – 104 °F)
Operating Altitude: Up to 2000 m (6500 ft)
Non-operating altitude: Up to 4600 m (14950 ft)
Safety standards: IEC,
CSA, UL: Installation Category II, Pollution Degree 2

Features

1. Up to 100 mL/min flow is provided at 400 bar without changing the pump head.
2. For preparation and analytical applications at low flow rates (0.5 mL/min, upward), providing an accurate high-pressure binary gradient structure with low delay volume (700 µL).
3. Extended seal life due to the use of peristaltic pump automatic cleaning seal.
4. Pump stops responding to column blockage or leakage.
5. Automatic solenoid cleaning valve system setup becomes simple.
6. The software provides solvent level control to keep the pump running dry.

Related Products

Waters 515 HPLC Pump, Waters Corporation(CAT#-STEM-C-0109-LGZ)
Thermo Electron Accela 600 Pump HPLC Quaternary Pump, Thermo Fisher Scientific(CAT#-STEM-C-0128-LGZ)
Persee L600 HPLC Pump, Persee(CAT#-STEM-C-0226-LGZ)
Shimadzu LC-20AT HPLC Pump, Shimadzu Excellence in Science(CAT#-STEM-C-0229-LGZ)
Pharmacia P-1 Peristaltic Pump, Pharmacia Fine Chemicals(CAT#-STEM-C-0251-LGZ)
Rainin Pressure Module Pressure Module, Rainin(CAT#-STEM-C-0258-LGZ)