Agilent 1100 Series - G1364A HPLC Fraction Collector, Agilent Technologies (CAT#: STEM-C-0047-LGZ)

Highlights

1. Agilent 1100 Series G1364A AFC Automatic distillate collector is used for scale preparation applications. Discrete distillate is one of its three operating modes, which is the default mode for all containers.
2. In the discrete fraction mode, the flow is diverted to the waste while moving from one container location to the next.
3. Continuous low should only be used when using the deep well plates.

Cat Number: STEM-C-0047-LGZ

Application: For preparative scale applications.

Model: 1100 Series - G1364A

Add to Cart



Description

The manufacturer of Agilent 1100 Series - G1364A HPLC Fraction Collector is Agilent Technologies. Model is 1100 Series - G1364A. Agilent 1100 Series G1364A AFC Automatic distillate collector is used for scale preparation applications. Discrete distillate is one of its three modes of operation, which is the default mode for all containers.

Specification

Condition: used
Trigger Modes: Time slices, Peak (threshold, up- /downslope), Timetable (combination of time intervals and peak) and Manual trigger (supported only with G1323BControl Module)
Operating Modes: Discrete fractions: default mode for all vessels. The flow is diverted to waste, while moving from one vessel position to the next vessel position Continuous flow: optional, available only when using well plates. It is possible to move from one well plate position to the next one without diverting the flow into the well plate to waste
Test tube /plate sizes: Minimum 48 mm to 100 mm maximum
Maximum tube volume: ca. 45 mL
Maximum flow rate: 100 mL / min (depending on viscosity and generated back pressure, max. 6 bar at the diverter valve)
Delay volumes [μL]:
Fraction collector inlet to diverter valve: -500 (typical, depends on length of the tubing)
Diverter valve: ~15
Diverter valve to needle: ~110
Needle: -~5
Delay calibration Sensor: Single wavelength absorbance detector working at 654 nm, consisting of a LED and a photo diode
Diverter Valve: 3/2 Diverter valve with low internal volume (15 μL), switching time < 100 ms, maximum operating pressure 6 bar
Cooling: Optional (with additional G1330B), performance depending on ambient conditions and the volume of collected fractions
Maximum capacity: 3 fraction collectors in parallel plus one recovery fraction collector connected via 12-Position, 13-Port Selector valve
GLP features Early maintenance feedback (EMF), electronic records of maintenance and errors
Interfaces: Controller-area network (CAN). optional; LAN or external contacts interface RS232C, APG-remote (for remote start / stop signals to / from other modules) Interface to G1330A Thermostat CAN-DC-out for operation of Agilent approved external devices like valves
Safety features: Leak detection and safe leak handling, error detection and display, exhaust fan for fume extraction of hazardous vapors

Features

1. Agilent 1100 Series G1364A AFC Automatic distillate collector is used for scale preparation applications. Discrete distillate is one of its three operating modes, which is the default mode for all containers.
2. In the discrete fraction mode, the flow is diverted to the waste while moving from one container location to the next.
3. Continuous low should only be used when using the deep well plates.
4. Needle Into Location where the needle pushes into the vessel as deep as specified, for the use with capped vials and test tubes and well plates with closing mats.
5. G1364A Thermostatted Fraction Collector, preparing scale, can be created by ordering additional and installing G1330A constant temperature distillate collector.

Related Products

Agilent 1100 Series - G1364C HPLC Fraction Collector, Agilent Technologies(CAT#-STEM-C-0048-LGZ)
Agilent 1100 Series - G1364D HPLC Fraction Collector, Agilent Technologies(CAT#-STEM-C-0049-LGZ)
Gilson FC204 HPLC Fraction Collector, Gilson(CAT#-STEM-C-0197-LGZ)
Transgenomic FCW 200 Wave HPLC Fraction Collector, Transgenomic (CAT#-STEM-C-0202-LGZ)
Advertisement