Condition: certified pre-owned
Dimensions: 28 × 31 × 16 cm: Accommodates up to two 105 m × 0.530 mm id capillary columns or two 10-ft glass
packed columns (9 in coil diameter, 1/4 in od), or two 20-ft stainless steel packed columns (1/8 in od)
Operating temperature range: Ambient temperature +4 °C to 450 °C
Temperature Ramps: Supports 6 oven ramps with 7 plateaus. Negative ramps are allowed
Maximum achievable temperature ramp rate: 120 °C/min (120 V units are limited to 75 °C/min
Maximum run time: 999.99 min (16.7 h)
Valve Support: Support for up to 8 valves.
– Valves 1-4, 12 V DC 13 watt in a heated valve box
– Valves 5-6, 24 V DC 100 mA unheated, for low power valve applications
– Valves 7–8, externally powered as a remote event from separate contact closure
Independent heated zones: Not including oven: Six (two inlets, two detectors, and two auxiliary)
Electronic Pneumatics Control (EPC):
• Atmospheric pressure compensation sensor to compensate for altitude and ambient temperature variation is standard.
• Pressure may be adjusted by increments of 0.01 psi.
• Pressure/Flow ramps: Three maximum.
• Carrier and makeup gas settings selectable for He, H2, N2, and argon/methane.
• Flow or pressure setpoints for each inlet or detector parameter with on both Agilent 6890N and the ChemStations.
• Automated carrier has constant flow, when capillary column dimensions are entered into the 6890.
• Split/Splitless, PTV inlets have flow sensors for the control of split ratio.
• Inlet modules Pressure sensors: Accuracy:
<±2% full scale, Repeatability:
<±0.05 psi, Temperature coefficient
: <±0.01 psi/°C, Drift:
<±0.1 psi/6 months.
Flow sensors: Accuracy: <±5%
depending on carrier gas,
Repeatability: <±0.35% of setpoint,
Temperature coefficient :
<±0.20 mL/min normalized temperature
and pressure (NTP)*
per °C for He or H2;
<±0.05 mL/min NTP per °C for
N2 or Ar/CH4.
• Detector modules
Accuracy: <±3 mL/min NTP or
7% of setpoint, Repeatability:
<±0.35% of setpoint, Temperature
coefficient: <±0.20 mL/min NTP
per °C
Inlets:
• Maximum of two inlets installed.
• Inlets available:
– Packed purged injection port
(PPIP)
– Split/Splitless capillary inlet
(S/SL)
– Temperature-programmable cool on-column (PCOC)
– Programmable temperature vaporizer (PTV)
– Volatiles inlet (VI)
S/SL:
• Suitable for all capillary columns.
• Split ratios up to 7500:1 to avoid column overload.
• Splitless mode for trace analysis. Pressure-pulsed splitless is
easily accessible for best performance.
• Maximum temperature: 400 °C.
• EPC available in two pressure ranges: 0–100 psig (0 to 680 kPa) for best control for columns 0.200 mm diameter; 0–
150 psig for columns <0.200 mm diameter. Pressure set point resolution is 0.01 psig. Total flow control/ sensing up to
1000 mL/min (He, H2).
• Gas saver mode to reduce gas consumption without compromising performance.
• Optimized septum purge flow to eliminate “ghost” peaks.
• Total flow setting range:
0 to 200 mL/min N2
0 to 1,000 mL/min H2 or He
FID Detector:
• Minimum detectable level (for tridecane): <1.8 pg C/s.
• Linear dynamic range: >107 (±10%). Digital data makes entire range available without range changes.
• Data rates up to 200 Hz accommodate peaks as narrow as 25 ms at half height.
• Standard EPC for three gases:
– Air: 0 to 800 mL/min
– H2: 0 to 100 mL/min
– Makeup gas (N2 or He): 0 to
100 mL/min
• Available in two versions: capillary column optimized or adaptable for either packed or capillary columns.
• Flame out detection and automatic reignition.
• Grounded jet.
• 450 °C maximum operating temperature.
TCD Detector:
• Minimum detectable level: 400 pg propane/mL with He carrier. (This value may be affected by laboratory
environment)
• Linear dynamic range: >105 ±5%.
• Unique fluidic switching design provides rapid stabilization from turn-on, low-drift performance.
• Signal polarity can be run programmed for components having higher thermal conductivity than the carrier gas.
• Maximum temperature: 400 °C.
• Standard EPC for 2 gases (He, H2, Ar, or N2 – matched to carrier gas type).
• Makeup gas: 0 to 12 mL/min.
• Reference gas: 0 to 100 mL/min
Micro-ECD Detector:
• Minimum detectable level: <6 fg/mL lindane.
• Proprietary signal linearization Linear dynamic range: >5 × 104 with lindane.
• Data acquisition rate: up to 50 Hz.
• Uses ß emission of <15 mCi 63Ni as the electron source.
• Unique micro-cell design minimizes contamination and optimizes sensitivity.
• 400 °C maximum operating temperature.
• Standard EPC makeup gas types: argon/5% methane or nitrogen; 0 to 150 mL/min.
• Dynamic range: >5 × 105 with lindane