The Atreus® Whole Blood Processing System is an independent automated manufacturing system that can process blood components of whole blood units. The system includes Atreus equipment, Atreus System Manager (ASM) and disposable processing set. This unique system simplifies whole blood processing by combining many labor-intensive manual processing steps with automation, thereby eliminating the variability from whole blood production and making your blood center more efficient and successful.
Specification
Size (W x H x D): 727 x 965 x 663 mm (29 x 38 x 26 in) Weight: 256 kg (565 lb) Humidity :5 to 80% RH non-condensing Operating temperature: 18 to 28°C Storage temperature: 0 to 60°C Pollution degree: 2 Installation category: II Environment: Indoor use only Rated voltage: 208-240 V, 50/60 Hz Rated input: 3.5 kVA Average input: 2.0 kVA Heat Output: 3412 Btu per machine Electrical supply: 20A, single phase Power plug: NEMA L6-20p Ventilation: Adequate to dissipate 1.0 KW in normal operation Minimum clearance envelope (perimeter around the Atreus machine): 152 mm (6 in) Minimum floor area needed for installation: 982 x 812 mm (39 x 32 in) Accuracy of volume measurement: ± 10% of predicted or within 10mL, whichever is greater Accuracy of time measurement: ± 5 sec. Rotor speed: 0 to 3,200 RPM Floor slope: 1.5° max Safety standards: IEC 60601-1, IEC 61010-2-20 Fluid leak detector: - Capable of detecting 20 mL/min in the rotor - Detects fluid leaks where moisture reaches the wall of the basin Noise level: <65 dB (SPL)