Conductivity measurement in Chemical Processing (CAT#: STEM-PPA-0021-WXH)

Introduction

Conductivity sensors and meters are used in chemical processing applications to monitor the conductivity of various solutions and mixtures. Conductivity measurements can be used to determine the concentration of different compounds in the solution and ensure that the process is running correctly.




Principle

Conductivity is the measure of the ability of any solution to pass an electrical current. Conductivity is usually affected by free ions which are dissolved in any solution. The ions can be chloride, sulfate, nitrate, phosphate, sodium, magnesium, iron, calcium, or aluminum.

Applications

Chemical Processing

Procedure

1. Use a conductivity standard solution (usually potassium chloride or sodium chloride) to calibrate the meter for the range that you will be measuring.
2. Rinse the probe with distilled or deionized water.
3. Select the appropriate range beginning with the highest range and working down. Read the conductivity of the water sample. If the reading is in the lower 10 percent of the range, switch to the next lower range. If the conductivity of the sample ex ceeds the range of the instrument, you may dilute the sample. Be sure to perform the dilution according to the manufacturer's directions because the dilution might not have a simple linear relationship to the conductivity.
4. Rinse the probe with distilled or deionized water and repeat step 3 until finished.

Materials

Conductivity Meters