Unlock Exclusive Discounts & Flash Sales! Click Here to Join the Deals on Every Wednesday!

Drager Air-Shields RW82-1 Resuscitaire Infant Incubator, Dräger (CAT#: STEM-RAFAEPO-0048-ZJF)

Highlights

Exceptional access from either side.
Configure to your department's needs.
Easily access suction.

Cat Number: STEM-RAFAEPO-0048-ZJF

Application: To provide controlled environment within which an infant can be cared for and kept warm

Model: RW82-1

Add to Cart



Description

The manufacturer of the Drager Air-Shields RW82-1 Resuscitaire Infant Incubatoris Dräger.
The Drager Air-Shields Resuscitaire RW infant warmer combines an effective warming therapy platform along with the components you need for clinical emergency resuscitation. The Drager Resuscitaire offers a blender, suction, flowmeter, and positive pressure mounted on the bed for easy accessibility. It features a Silica Quartz Heater that provides complete, even and quick heating. A parabolic reflector helps distribute the heat evenly over the entire mattress area.
The product is located in Big Lake, MN, USA.

Specification

Condition: Certified Pre-owned
Physical Attributes (Without Options / Accessories)
Height: 72 in (183 cm)
Width: 26.5 in (68 cm)
Depth: 44 in (112 cm)
Weight: (without options /accessories) 220–280 lbs (100–127 kg)
Mattress
Height From Floor: 39.4 in (100 cm)
Width: 21.0 in (53.3 cm)
Length: 26.0 in (66.0 cm)
Thickness: 1.0 in (2.54 cm)
Maximum Capacity: 15 lbs (4.54 kg)
Maximum Power Requirements
Model 120V: 120 V, 50–60 Hz, 750 W
Model 220/240V: 220–240 V, 50–60 Hz, 750 W )
LED Examination Light
Illumination: > 0.11 lumens / cm2 (100 ft. Candles)
Controller Heating Alarms
Check patient System alerts for 5 minutes in Manual Mode every 30 seconds after 10 minutes
Baby temperature ±1° C from Set Point
High temperature Skin Temp. 39.0 ±0.2 °C
Probe Short or open circuit / No probe
System fail Indicates System Fail
Power fail AC Power Interruption
Alarm Silence / Reset Intervals
Check patient Resets clock for 10 minutes manual mode
Baby temperature 10 minutes
High temperature 2 minutes
Procedural silence Presilences baby temp alarm for 5 minutes
Set Temperature / Heater Output
Set Temp Range: 34–38 °C (>37.0° override)
Heater Output: 0–100%, 10% increments
Skin Temp Display: 18–43 °C
Accuracy: ±0.2 °C (31 to 37° C)
Resuscitation Module
Wall Supply Pressure: 40–75 psi (275–517 kPa)
Cylinder Pressure: 2,900 psi max (19 994 kPa)
Patient Gas Supply
Flow Control Range: 0–15 LPM
Airway Pressure Limit: Adjustable 0–50 cm H2O (0–4.9 kPa); Internal preset limit 60 cm H2O (5.9 kPa)

Features

Exceptional access from either side.
Configure to your department's needs.
Easily access suction.
Efficiently adjust mattress position.
Head of the warmer module can swivel up to 90 degrees.
Automatic Pre-Warm Mode allows long-term, controlled warming of the mattress with no nuisance alarms.

Related Products

Drager Infinity Gamma XL Multiparameter Patient Monitor, Dräger
Drager Vapor 19.3 Isoflurane Vaporizer, Dräger
Drager Vapor 19.1 Isoflurane Vaporizer, Dräger

Other recommended products

Advertisement