Maximum weight (g): 220g
Readability (mg): 0.01mg
Repeatability sd (mg)-20g: 0.015mg
Repeatability sd (mg)-100g: 0.03mg
Linearity error (mg): ±0.1mg
Span calibration point (g): 100g; 150g; 200g
Calibration: AutoCal™ automatic calibration every Δ1.5˚C temperature difference or 3 hour intervals
Automatic hood door: Standard
Weighing units: milligrams, grams, carats, ounces, troy ounces, bucks, grains, Momme, Mesghal, Tical, Tola, Baht, 3 custom units
Weighing modes: basic weighing, piece counting, percentage weighing, check weighing, animal weighing, target value weighing, cumulative weighing, proportioning weighing, differential weighing, density determination, peak hold, shift Liquid vessel calibration, statistical quality control SQC
Stabilization time (seconds): 0.1mg: ≤3 seconds, 0.01mg: ≤8 seconds
Temperature Drift: 0.50 ppm /˚C
Typical minimum weight: 20mg (USP,>
Best Minimum Weight: 9mg (USP,>
Display type: 5.7-inch VGA color touch display / 14.5 cm (diagonal)
Communication interface: Standard: 2 USB, 1 RS232; Optional: 2nd RS232 interface, Ethernet interface
Operating Environment: 10˚C – 30˚C
Storage Conditions: -10˚C to 60˚C, 10% to 90% relative humidity, non-condensing
Power: Input Power: 100-240 VAC 0.6A 50-60Hz; Adapter Output Power: 12VDC 1.5A
Weighing pan size: Ø 80 mm
Overall Dimensions (W x H x D): 230 x 393 x 350 mm
Shipping Dimensions (W x H x D): 632 x 640 x 467 mm
Net weight: 7.5kg
Shipping weight: 12.6kg