Light source: Deuterium lamp
Wavelength range:
190 ~ 600 nm
Cut-off filter to eliminate second-order diffraction is automatically activated for the 371- 600nm range
Spectral bandwidth: 8 nm
Wavelength accuracy: ± 1 nm
Wavelength reproducibility: ± 0.1 nm
Cell path length: 10nm
Cell volume: 8 µL
Cell pressure tolerance: 12MPa
Dual wavelength mode:
Measurement wavelength: Selectable, two wavelengths in 190- 370 nm or 371-600 nm
Sampling frequency: 1.2 sec for one wavelength
Response: Selectable in 10 steps corresponding to time constant 0.05, 0.5, 1.0, 1.5, 1.0, 1.5, 2.0, 3.0, 6.0, 8.0, 10.0 seconds
Range: Can be set between 0.0001 and 2.56 AUFS in 0.0001 AUFS steps
Zero adjustment: Auto zero function, baseline shift function
Wavelength scanning function: Performs wavelength scanning with flow stopped. Data stored in three files, one of which stores background. Background is subtracted from sample scans. File data is not saved up (lost when power is turned OFF).
Time Program: Available in detector or by system controller
Set items:
Wavelength (including dual wavelengths), auto zero, range, marker, response, wavelength scanning, event, polarity, lamp ON/OFF, loop
Number of steps: Maximum 32 steps