Condition: Certified pre-owned
Read modes:
Absorbance
Fluorescence Intensity
Luminescence
Time-Resolved Fluorescence
Fluorescence Polarization
Robotic-compatible: Yes
Plate formats: 6, 12, 24, 48, 96, 384 wells
Light source: Xenon Flash Lamp (1 joule/flash)
Detectors: 2 photomultiplier tubes (PMT)
Shaker time: 0 to 999 seconds
Temp. control: 2°C above ambient to 60°C
Temp. uniformity: < 1°C at 37°C set point
Temp. accuracy: ±1°C at 37°C set point
Readings: Endpoint, Kinetic Spectral Scanning
Standard Read Times (minutes: seconds) 96 wells 384 wells:
Absorbance 0: 18 0: 49
Fluorescence Intensity 0: 17 0: 48
Luminescence 2: 00 7: 00
Absorbance Photometric Performance:
Reading Capabilities: Cuvette or microplate
Wavelength range: 200–1000 nm
Wavelength selection: Monochromator, tunable in 1.0 nm increments
Wavelength bandwidth: ≤ 4.0 nm
Wavelength accuracy: ±2.0 nm
Wavelength repeatability: ±0.2 nm
Photometric range: 0–4.0 OD
Photometric resolution: 0.001 OD
Photometric accuracy:
(microplate) < ±0.006 OD ±1.0%, 0–2 OD
(cuvette) < ±0.005 OD ±1.0%, 0–2 OD
Photometric precision: < ±0.003 OD ±1.0%, 0–2 OD
Stray light: < 0.05% @ 230 nm
Fluorescence Intensity Performance: Top or bottom
Wavelength range: 250–850 nm
Wavelength selection: Monochromators, tunable in 1.0 nm increments
Bandwidth (EX, EM): 9 nm, 15 nm
Optimized sensitivity:
≤ 1 pM fluorescein in 96 wells,
≤ 1.5pM in 384 wells
Guaranteed sensitivity:
< 5 pM fluorescein in 96 wells or cuvette,
< 20 pM in 384 wells
Luminescence Performance: Top or Bottom
Wavelength selection: Choice of simultaneous detection of all wavelengths or selection via monochromator, tunable in 1.0 nm increments
Wavelength range: 250–850 nm
Optimized sensitivity: ≤ 43pM ATP in 96 wells
Guaranteed sensitivity: ≤ 75pM ATP in 96 wells
Dynamic range: > 6 decades
Cross-talk: < 0.3% in white 96- and 384-well Microplates
Time Resolved Fluorecesnce Performance: Top or Bottom of a Microplate
Wavelength selection Monochromators,: tunable in 1.0 nm increments
Bandwidth (EX, EM): 9 nm, 15 nm
Precision data collection: 1–100 flashes, delay of 0–600 µsec. before read, integration time selectable between 50–1500 µsec.
Optimized sensitivity: ≤ 10 fM europium in 96
Guaranteed sensitivity: ≤ 100 fM europium in 96 or 384 wells
Fluorescence Polarization Performance:
Wavelength range: 300–750 nm
Wavelength selection: Monochromators, tunable in 1.0 nm increments
Bandwidth (EX, EM): 9 nm, 15 nm
Optimized Precision: ≤ 3.5 mP standard deviation at 1 nM fluorescein in 96 wells
Guaranteed Precision: < 5 mP standard deviation at 1 nM fluorescein in 96 wells
Dimensions: 22x16x9"
Weight: 36 lbs
Power: < 420 watts, 100–240 VAC, 3.5 A, 50/60Hz