Unlock Exclusive Discounts & Flash Sales! Click Here to Join the Deals on Every Wednesday!
Sterilization is defined as using steam, dry heat, hydrogen peroxide, ethylene oxide, or radiation to render a device free from viable microorganisms. The level of sterilization is defined by Sterility Assurance Level (SAL), which is the probability that a device is not sterile. For example, an SAL of 10-6 indicates a 1 in 1 million possibility that the device is non-sterile.
Sterilization Validation is used to help manufacturer establish appropriate sterilization parameters for health care processing instructions, and confirm the pertinent SAL, dry time and thermal profiling of the medical device. For non-critical devices (those that do not penetrate the body), the sterilization process is considered to be validated if the SAL is 10-3, while for critical devices, the SAL should be 10-6. Based on the data obtained through testing, the manufacturer will be able to provide health care facilities with validated sterilization parameters.
STEMart helps manufacturer test and validate the labeling claims for sterilization of reusable medical devices in a health care facility.
STEMart provides adequate testing to demonstrate the efficacy of recommended sterilization instruction. If you want to learn more detail about our service, or would like to consult with the experts at STEMart, please feel free to contact us.