Dimensions
Height: 11.4 In (290 cm)
Width: 20.9 In (530 cm)
Depth: 17.7 In (450 cm)
Weight: Approx. 27 kg (60 lbs)
Diagonal Screen size: 6.5" TFT LCD color Display
Ventilation Settings
Ventilation mode: IPPV, IPPVAssist (CMV, CMVAssist) SIMV, SIMVASB (SIMV, SIMV/Psupp) MMV, MMVASB (MMV, MMV/Psupp) BIpaP1), BIpaP1)ASB, BIpaP1)Assist (PCV+, PCV+/Psupp, PCV+Assist) CPAP, CPAPASB (CPAP, CPAP/Psupp) APRV (optional) ILV (optional)
Enhancements: AutoFlow™ – Automatic adaptation of inspiratory flow in volume controlled modes ATC™ – Automatic Tube Compensation (optional) IV - Mask Ventilation (optional)
Ventilation frequency (f): 0 to 100 /min, 0 to150 /min (Neonatal)
Inspiration time (Tinsp): 0.1 to 10 s
Tidal volume (VT) (BTPS): 0.1 to 2.0 L (Adult) 0.02 to 0.3 L (Pediatric) 0.003 to 0.1 L (Neonatal)
Inspiratory flow: 6 to 120 L/min (Adult) 6 to 30 L/min (Pediatric and Neonatal)
Inspiratory pressure: 0 to 80 mbar (cmH2O) PEEP / intermittent PEEP: 0 to 35 mbar (cmH2O)
Pressureassist (PASB) (Psupp): 0 to 80 mbar (cmH2O)
Rise time for inspiratory pressure: 0 to 2 s
O2 concentration: 21 to 100 Vol.%
Multi-sense Trigger Criteria: Internal automatic pressure trigger, Flow, Volume (Flow adjustable 0.3 to 15 L/min)
Measured Values Displayed
IAirway pressure: Peak pressure, plateau pressure, mean pressure, PEEP, min. pressure (0 to 99 mbar/cmH2O)
Minute volume (MV), (BTPS): MV, MVspont, MVleak (0 to 99 L/min)
Tidal volume (VT), (BTPS): Inspired VT, expired VT, VTASB (VTPS) (0 to 3999 mL)
Breathing frequency (f): ftotal, fspon, fmand. (0 to 300/bpm)
O2 concentration: (FiO2) Inspired O2 concentration (15 to 100 Vol.%)
Lung mechanics: Resistance ((0.0 to 600 mbar/L/s)(cmH2O/L/s) Compliance ((0.0 to 300 mL/mbar)(mL/cmH2O)
Breathing gas temperature: 18°to 51°C
Waveforms: Airway pressure-time, flow-time, volume-time, ...
Trends (optional): FiO2, MV, VT, f, PEEPi, R, C, etCO2, ...
Loops (optional): Paw-V, V-Flow, Flow-Paw, ...
Capnography (etCO2) (optional): 0 to 100 mmHg
CO2 production (VCO2): 0 to 999 mL/min, STPD
Serial dead space Vds: 0 to 999 mL, BTPS
Alarms / Monitoring
Airway pressure: High/Low
Expired minute volume: High/Low
Tidal volume: High
Tidal volume (VT) (BTPS): 0.1 to 2.0 L (Adult) 0.02 to 0.3 L (Pediatric) 0.003 to 0.1 L (Neonatal)
Inspiratory flow: 6 to 120 L/min (Adult) 6 to 30 L/min (Pediatric and Neonatal)
Apnea alarm time: 5 to 60 s
Spontaneous breath frequency: High
Inspired O2 concentration: High/Low
Breathing gas temperature: High
SpO2 pulse (optional): High/Low
etCO2 (optional): High/Low
Performance Data
Max. flow for pressure support and spontaneous breathing: 180 L/min (adult), 60 L/min (pediatric)
Valve response time: ≤ 5 ms
Control principle: Time cycled, volume constant, pressure-controlled
Safety relief valve: 100 mbar (cmH2O)
Leakage compensation
Hose system compensation
Outlet for pneumatic nebulizer