●Background Count: ≤60CPM
● Detection Efficiency for I-125: ≥78%
8 hours stability: the additional error of detection efficiency is not more than 3%
●Counting Precision: Repeat the measurement 21 times with the X2 double test method, and meet 10.851≤X2≤31.410
●Number of Probes: 1
●One-time Loading Capacity: 250 samples
●Instrument Working Mode: automatic measurement
Function
●Multiple radionuclide markers such as I-125, Co-57, Fe-59, Se-75, ICo-I can be selected for measurement.
●Automatic alignment measurement. Using barcode reading automatic identification technology, up to 10 detection items can be set at one time.
●It has the function of continuous measurement after power failure, concentration reverse check function, standard curve calling function, etc.
●It has the function of double-pipe measurement. The maximum number of multiple tubes can be set to 4 to improve the detection accuracy.
●The instrument has a high degree of intelligence, and can automatically measure the plateau and spectral curve, fault prompts, etc., and the operation and maintenance are simple
●Six Detection Modes to Choose: radioimmunoassay (RIA), immunoradiation (IRMA), hepatitis, TG.TM, insulin series, pure count measurement
●Seven Data Processing Methods: linear interpolation equation, spline culvert, cubic polynomial equation, Log_logit equation, 3/2 equation, four-parameter, five-parameter
●Five Binding Rate Modes: B/T, B/B0, B/Bm, F/B, B/F
●Can edit and print a variety of report styles.
●Quality control analysis function: provide inter-batch quality control, intra-batch quality control, precision chart, quality control chart, etc.
●Data can be transmitted online to realize resource sharing