Overall Performance
Throughput: 200Tests/Hour
Analysis Method: End-point, Fixed-time, Rate(Kinetic), Turbidimetry
Sample & Reagent Unit
Sample Positions: 49 sample positions
Reagent Positions: 56 reagent positions
Sample Volume: 2~70μl
Reagent Volume: 20~350μl
Probewashing: Automatic washing interior and exterior
Reagent Cooling: Refrigerated tray with independent switch
Reaction System
Temperature Control: 37±0.1°C, real-time monitor
Cuvettes: 120 reusable cuvettes, optical length 6mm
Mixer Probe: Independent stirring
Washing: Automatic cuvettes washing
STAT function: YES
Optical System
Light Source: 12V/30W halogen lamp
Spectrophotometry: Post-spectral spectrophotometry
Wavelength: 340,405,450,510,546,578,630,700nm
Absorbance: 0~3.0Abs
Calibration & QC
Calibration: Linear: K factor, 1-point, 2-point and multipoint linear; Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P
Quality Control: Real-time QC, Westgard multi rule, Cumulative sum check, Twin Plot(2D)
Data Management
Software: Windows 7/8/10, LIS system available
LIS System: Bi-direction, support HL7 protocol
Interface: LAN port access
Printer: External, multiple reporting mode available
Working Conditions
Power Supply: AC220V±10%, 60/50Hz, 110V±10%, 60Hz, 300W
Temperature: 10~30°C
Water Consumption: Deionized water: 5L/H
Humidity: 30-80%
Size & Weight
External Size (W*D*H): 950*612*510mm
Net Weight: 75kg