Overall Performance
Throughput: 400 Test/hour
Analysis Method: End-point, Fixed-time, Rate(Kinetic), Turbidimetry
Sample & Reagent System
Sample Positions: 60 or 90 sample positions, built-in barcode system available
Reagent Positions: 45 refrigerated reagent positions
Sample Volume: 2~70μl
Reagent Volume: 20~350μl
Probe: Teflon coating, with anti-collision function, liquid level detection function
Probe Washing: Automatic washing interior and exterior
Reagent Cooling: Refrigerated tray with independent switch
Reaction System
Temperature Control: 37±0.1°C, real-time monitoring
Cuvettes: 120reusable cuvettes, optical length 6mm
Mixer Probe: Independent stirring
Washing: Automatic cuvettes washing
STAT function: YES
Optical System
Light Source: 12V/30W halogen-tungsten lamp
Spectrophotometry: Post-spectral spectrophotometry
Wavelength: 340,405,450,510,546,578,630,700nm
Absorbance: 0~3.0Abs
Calibration & QC
Calibration: Linear: K factor, 1-point, 2-point and multipoint linear; Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P
Quality Control: Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D)
Data Management
Software: Windows 7/8/10, 32or 64bit
LIS System: Bi-direction, support HL7 protocol
Interface: LAN port access
Printer: External, multiple reporting mode available
Working Conditions
Power Supply: AC220V±10%, 50/60Hz, 110V±10%, 60Hz, 300W
Temperature: 10~30°C
Water Consumption: Deionized water: 12L/H
Humidity: 30-80%
Size & Weight
External Size (W*D*H): 1167*780*1016mm
Net Weight: 193kg
Package Size (W*D*H): 1348*928*1399mm
Gross Weight: 267kg