Culture room temperature range: TS/TSD: ambient temperature 15℃ ~ 30℃, temperature control range: RT+3℃ ~ 60℃, TSZ: ambient temperature 15℃ ~ 30℃, temperature control range: 10℃ ~ 60℃
Temperature uniformity: ≤±1℃
Temperature fluctuation: ≤±0.5℃
Anaerobic grade: Oxygen content in operation room ≤0.1%
Oxygen concentration display range: 0 ~ 21%
Carbon dioxide concentration display range: 0 ~ 20%
Humidity display range: 0 to 100%
Time of anaerobic formation in sampling chamber: ≤15 minutes
Anaerobic method of sampling chamber formation: vacuum + gas displacement type (99.99% nitrogen or hydrogen 5% carbon dioxide 10% nitrogen 85% mixture)
Sampling chamber to form micro-aerobic mode: vacuum + gas displacement type (99.99% nitrogen or oxygen 5% carbon dioxide 10% nitrogen 85% mixture)
Operation room forming anaerobic state time: ≤1.5 hours operation room oxygen content ≤0.3%
Operation chamber formed anaerobic mode: gas displacement type (99.99% nitrogen or hydrogen 5% carbon dioxide 10% nitrogen 85% mixture) + micro flow control
The operation room forms micro-aerobic mode: gas displacement type (99.99% nitrogen or oxygen 5% carbon dioxide 10% nitrogen 85% mixture) + micro-flow control
Operation room anaerobic environment maintenance: automatic
Control mode of anaerobic formation: automatic execution with one key
Sterilization method: ultraviolet sterilization lamp ×1
Sampling chamber size W*D*H(cm): 40×33×32
Operation room size W*D*H(cm): 91×65×60
Culture chamber size W*D*H(cm): TS/TSD: 42×30×50
External dimension W*D*H(cm)
TS/TSZ: 140×73×137, TSD: 140×73×70
Packing size W*D*H(cm): 154×87×155
Display interface: 7 "color touch screen P.I.D. controller, Chinese and English display
Power supply: 1Ø 220V 50Hz
Gas supply requirements: gas purity 99.99%, 0.05mpa, continuous gas supply
Ambient temperature: Available ambient temperature: 5 ℃ to 35 ℃ Optimal ambient temperature: 15 ℃ to 30 ℃
Type III: standard pressure sensor, oxygen sensor, temperature sensor, the value of the same screen display; Type A: standard with pressure sensor, oxygen sensor, carbon dioxide sensor, temperature sensor, humidity sensor, numerical display with the same screen, standard with USB data recording serial port; Support control program USB automatic upgrade; Type A2: standard pressure sensor, oxygen sensor, carbon dioxide sensor, temperature sensor, humidity sensor, the value of the same screen display; Independent carbon dioxide concentration control, standard USB data recording serial port; Support control program USB automatic upgrade, standard wireless, wired remote alarm module, remote monitoring through mobile APP
Note: TS: Culture room without refrigeration system; TSZ: No culture room, operation room overall temperature control, refrigeration system; TSD: table top (no bracket), culture room, no refrigeration system