Solvent Manager:
Number of solvents: Up to four, in any combination of two: A1 or A2 and B1 or B2.
Solvent conditioning: Vacuum degassing (six-channel): one channel per solvent, and two channels for sample manager wash solvents.
Settable flow rate range: 0.001 to 2.000 mL/min, in 0.001 mL increments
Gradient profiles: Eleven gradient curves Wet prime Automatic
Maximum operating pressure: 103,421 kPa (1034 bar, 15,000 psi) up to 1 L/min, 62,053 kPa (621 bar, 9000 psi) up to 2 mL/min
Composition accuracy: ±0.5% absolute (full scale) from 5 to 95% from 0.50 to 2.00 mL/mina
Composition precision: 0.15% RSD or ±0.04 min SD
Flow accuracy: ±1.0% at 0.5 mL/min
Sample Manager:
Number of SBS/ANSI sample plates:
Total of two plates, expandable to up to 22 plates with optional sample organizer.
Maximum of 22 microtiter plates, 11 intermediate-depth plates, or 8 deep-well (or 2 mL vial) plates:
96 and 384 microtiter plates
48-position 2 mL vial plates
48-position 0.65 mL micro-centrifuge tube plates
24-position 1.5 mL micro-centrifuge tube plates
Maximum sample capacity: 768 in two 384-well plates; expandable to up to 8,448 samples with optional sample organizer
Injection volume range:
0.1 to 50 µL, in 0.1 µL increments,
partial or full loop mode, 10 µL loop is
standard; 1, 2, 5, 20, and 50 µL loops
Sample temperature control: 4 to 40 °C, programmable in 0.1 °C increments
Minimum sample required: 5 µL residual volume at a needle offset of 0, or 22 µL at the preset 2 mm needle offset
Column Heater:
Columns accepted: Any up to 4.6 mm ID and 150mm long
Temperature range: 5 above ambient to 65°C (9 above ambient to 194°F)
PDA Detector:
Wavelength Range: 190-500 nm
Flow Cell Options:
Analytical Cell 500 nanoliters, 10 mm pathlength
High Sensitivity Cell 2.4 microliters, 25mm pathlength
Dimensions: 25 x 14 x 34”
Power: 100-240v, 50/60 Hz
Includes:
Waters Acquity UPLC System with TUV & Column Manager
Waters Acquity Tray
Waters Acquity TUV Detector
Waters Acquity Column Manager
Waters Acquity Sample Manager
Waters Acquity Binary Solvent Manager
Data System