Throughput: 60Tests/hour
Measuring Principle: Impedance for cell counting Cyanide-free method for HGB
Aperture Diameter: WBC chamber: 100μm, RBC/PLT chamber:70μm
Sample Volume: Normal: 10μl; Predileted: 20μl
Calibration: Manual & auto calibration
Counting Modes: Venous, capilary, prediluted
Printer: Built-in thermal printer, USB to external printer(optional)
Data Input: 8 inch touch screen, keyboard, mouse
Output: RS232 port, USB port, VGA port. LIS system avaliable
Storage: 100,000 results with histogram
Power supply: AC 220±10%, 60/50Hz; 110±10%, 60Hz
Package Size: 580*470*590mm (W*D*H)
Gross Weight: 27kg
Carry-over Rate
WBC≤ 2%
RBC≤ 1%
HGB≤ 2%
PLT≤ 2%
Background
WBC≤ 0.2×10⁹/L
RBC≤ 0.02×10¹²/L
HGB≤ 1 g/L
HCT≤ 0.5%
PLT≤ 10×10⁹/L
Precision
WBC: CV≤ 2.0%, (4.0-15.0×10⁹)/L
RBC: CV≤ 1.9%, (3.5-6.0×10¹²)/L
HGB: CV≤ 1.9%, (110.0-180.0g)/L
MCV: CV≤ 0.4%, (80.0-110.0)fL
PLT: CV≤ 4.0%, (100.0-500.0×10⁹)/L
Linearity
WBC: CV: (±0.3×10⁹)/L, (0-5.9×10⁹)/L; CV: ±5%, (6.0-99.9×10⁹)/L
RBC: CV: ((±0.05×10¹²)/L, (0-0.99×10¹²)/L; CV: ±5%, (1.0-9.99×10¹²)/L
HGB: CV: (±2.0g)/L, (0-99.0g)/L; CV: ±2%, (100.0-300.0g)/L
PLT: CV: (±8.0×10⁹)/L, (0-99×10⁹)/L; CV: ±10%, (100-999×10⁹)/L