Condition: certified pre-owned
Analytical System: Fully automated, random-access clinical chemistry system with STAT capability
Analytical Principles: Spectrophotometry and potentiometry
Analytical Types: Endpoint, rate, fixed point and indirect ISE
Analytical Methods: Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE
Simultaneously Processed Analytes: 60 photometric tests + 3 ISE
Throughput: 800 photometric tests/hour; up to 1,200 with ISE
Sample Types: Serum, plasma, urine, whole blood (HbA1c) and other fluids
Sample Feeder: Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 150 samples; continuous loading
Sample Tubes: In primary and secondary tubes; diameter between 11.5 and 16.5 mm; height between 55 and 102 mm
STAT Samples: Up to 22 positions for STAT samples, barcoded primary tubes
Sample Volume: 1–25 μL in 0.1 μL increments (1–25 μL for repeats)
Reagent Supply: 60 positions for R1, 48 positions for R2; refrigerated: 4–12°C; handling of 3-shot reagents
Reagent Volume: R1: 15–250 μL; R2: 15–250 μL; in 1 μL increments
Total Reaction Volume: 120–425 μL
Reaction Cuvette: Quartz cuvettes
Reaction Time: Up to 8 minutes, 40 seconds
Reaction Temperature: 37°C
Mixing Method: With rotating paddles after dispensing sample and reagent
Photometry System: Direct assay through the reaction cuvette (0–3.0 OD) mono and bichromatic measurements possible
Wavelength: 13 different wavelengths between 340–800 nm
Cuvette Cleaning: Comprehensive cleaning with detergents
Calibration: Autocalibration, cooled calibrator positions; master calibration established by two-dimensional barcode on the reagent bottle (>3-point calibration)
Quality Control: Auto QC, cooled QC positions
Test Requisition: Individual and profile test requisition via online, mouse, function keys or touch screen
Safety: Clot detection and crash prevention for sample and reagent dispenser
Online: Full uni- and bi-directional communication possible
Software: Windows 7®
Data Storage: Sample number: 100,000 samples. Reaction data: 200,000 tests.
Dimensions (WxHxD): ANL 1,950×1,280×1,000
Power Supply: 200 V; 208 V; 220 V; 230 V; 240 V; 50 Hz; 60 Hz/3.8 KVA