OXYLETPRO - TREADMILL WITH INDIRECT CALORIMETRY, Panlab (CAT#: STEM-LEAR-1303-YJL)

Highlights

New control unit with integrated touchscreen graphic user interface
Adjustable belt speed up to 150 cm/s and belt slope from -25 to +25 degrees
Models mice and rats

Cat Number: STEM-LEAR-1303-YJL

Application: For assessing respiratory metabolism in treadmill studies

Model: OxyletPro

Add to Cart



Description

OXYLETPRO - TREADMILL WITH INDIRECT CALORIMETRY, model OxyletPro is manufactured by Panlab
The OxyletPro system is a modular system allowing the integration of respiratory metabolism (O2 consumption /CO2 production). Panlab provides a complete Oxylet solution enabling forced exercise training in a treadmill combined with indirect calorimetry in rodents.
The LE400 air flow control unit allows a fine regulation of the air flow inside the treadmill and sends the air to the LE450 gas analyzer for O2/CO2 gas concentrations determination.
The associated METABOLISM software transfers the data from the treadmill control unit and the gas analyzer to a PC computer using RS232/USB outputs for data storage and further analysis.
NOTE: the calibration tanks are not provided with the system and should be purchased separately by the user (see Specifications tab).

Specification

Treadmill
Current range: adjustable from 0 to 2 mA
Belt speed: adjustable from 0 to 150cms/sec
Running lanes: 1
Slope: adjustable from -25° to 25°
Exercise area dimensions (cm): 53 (l) x 10 (w) x 15 (h)
Treadmill dimensions - without lid (cm): 65 (l) x 17 (w) x 24 (h)
Base/Foot dimensions (cm): 40 (l) x 33 (w)
le405 oxygen sensor
Technology: Laser Diode Absorption
Measurement range: 2-100% (only limited by the high calibration point used)
Resolution: 0.01%
Linearity: +/- 0.2%
Noise: +/- 0.03% (20 ms average)
Accuracy: +/- 0.2% (24 hours)
 
le405 carbon dioxide sensor
Technology: Infrared Spectroscopy
Measurement range: 0-10%
Resolution: 0.01%
Accuracy: < 10% of reading between 5%-10% CO2; < 0.3% absolute for reading < 5% CO2
Room specifications for indirect calorimetry
Room temperature: 20-28 °C, the important is that the temperature is maintained stable throughout the experiment.
Ventilation: 10-20 room air exchanges per hour
Humidity: 30-70 %Rh, the important is that the humidity is maintained stable throughout the experiment.
Barometric pressure: The important is that the barometric pressure is maintained stable throughout the experiment.
Room access: Dedicated closed space with controlled and reduced access to the staff

Features

New control unit with integrated touchscreen graphic user interface
Adjustable belt speed up to 150 cm/s and belt slope from -25 to +25 degrees
Models mice and rats
Constant electrical shock intensity (adjustable from 0 to 2 mA)
Indirect calorimetry measurement for VOmax determination

Related Products

OXYLETPRO – WASTE COLLECTION, Panlab (CAT#: STEM-LEAR-1304-YJL)
ANESTHESIA Systems (Harvard Apparatus), Panlab (CAT#: STEM-LEAR-1305-YJL)
AUTO TAIL SUSPENSION SYSTEM, Panlab (CAT#: STEM-LEAR-1306-YJL)
CAGING (Panlab), 220 x 220 x 145 mm, Panlab (CAT#: STEM-LEAR-1307-YJL)
CAGING (Panlab), 215 x 465 x 145 mm, Panlab (CAT#: STEM-LEAR-1308-YJL)
CAGING (Panlab), 215 x 465 x 190 mm, Panlab (CAT#: STEM-LEAR-1309-YJL)
Advertisement