SKU: 626650
ProSolo Specifications
Dimensions: 8.3 cm width x 21.6 cm length x 5.6 cm depth; 567 g (with battery)
Power:
Rechargeable lithium-ion battery pack provides ~48 hours with the handheld only;
Battery recharge time is ~ 9 hours with the AC power adapter;
The instrument can also be powered via AC or external power through the USB port.
Operating Temperature: 0 to 50°C
Storage Temperature: 0 to 45°C with battery installed; 0 to 60°C without battery installed
Display: Color, LCD graphic display; 3.9 cm width x 6.5 cm height
Memory: >100,000 data sets
Barometer Units: mmHg, inHg, mbar, psi, kPa, atm:
Range: 375 to 825 mmHg;
Accuracy: ±1.5 mmHg from 0 to 50°C;
Resolution: 0.1 mmHg;
Sites and Data ID: 100 user-defined sites and 100 user-defined data ID tags; Site pictures can be sent to the handheld via KorDSS Software.
Calibration Records: 400 detailed calibration records can be stored and are available to view, download, and print (printing only available via KorDSS Software)
Languages: English, Spanish, German, French, Italian, Norwegian, Portuguese, Japanese, Chinese (Simplified & Traditional), Korean, Thai
Certifications: CEC, CE; RoHS; IP-67; WEEE; FCC; UN Part III, Section 38.3, Test methods for lithium-ion batteries (Class 9)
Warranty: NMEA 0183; requires user-supplied GPS and YSI 6115 Y-cable
ODO/CT Specifications
Size: 2.46 cm diameter, cable options for 1, 4, 10, 20, 30, 50, and 100 m lengths
Dissolved Oxygen (Optical luminescence) Units: % saturation, % saturation local, mg/L, ppm:
Range: 0 to 500%, 0 to 50 mg/L
Accuracy: 0 to 200%: ±1% of reading or 1% saturation, whichever is greater
200 to 500%: ±8% of reading
0 to 20 mg/L: ±0.1 mg/L or 1% of reading, whichever is greater
20 to 50 mg/L: ±8% of reading
Resolution: 0.01 mg/L and 0.1%, or 0.1 mg/L and 1% (auto-adjusts based on range)
Conductivity (Four nickel electrode cell) Units: μS/cm, mS/cm:
Range: 0 to 200 mS/cm
Accuracy: 0 to 100 mS/cm ±0.5% of reading or .001 mS/cm, whichever is greater
100 to 200 mS/cm ±1.0% of reading
Resolution: 0.001, 0.01, or 0.1 mS (range dependent)
Specific Conductance Units: μS/cm, mS/cm:
Range: 0 to 200 mS/cm
Accuracy: 0 to 100 mS/cm ±0.5% of reading or .001 mS/cm, whichever is greater
100 to 200 mS/cm ±1.0% of reading
User selectable reference temperature (15 to 25°C; default 25°C) and compensation coefficient (0 to 4%/°C; default 1.91%)
Resolution: 0.001, 0.01, 0.1 mS/cm
Warranty: 2 years on cable, probe, and ODO sensor cap