Light source: DC12V 22W KLS halogen tungsten lamp, with power management
Measurement channel: vertical 8 optical channels, automatically find the center
Wavelength range: 400 - 900nm
Optical filter: Standard 405 (412), 450, 492, 630, four standard optical filters, other wavelengths are optional, up to 10 optical filters can be loaded
Reading range: 0.000 - 4.0000Abs (405nm)
Linear range: 0.000 - 3.000Abs (405nm)
Repeatability: CV≤0.3%
Stability: ≤±0.003Abs
Indication error: ≤±0.01Abs
Resolution: 0.001Abs (display, print), internal calculation 0.0001Abs
Layout: visual layout, support standard 96-hole free layout
Vibration plate function: Vibration plate intensity (from weak to strong) 3 levels optional, vibration plate time 0-255 seconds adjustable, error ± 2 seconds
Display input: 8-inch color LCD display, touch screen input, visual layout, and display of the entire board information
Sampling: automatic opening/closing of the sampling chamber door, no need to manually open the cover
Workstation: professional microplate reader software with single and double wavelength, multi-wavelength and inhibition rate measurement functions, can store 500 sets of programs, 200,000 sample results, provide absorbance, threshold, linear equation, logarithmic equation, quadratic equation, cubic equation , exponential curve, power function, absorbance percentage-concentration logarithmic equation, 4-parameter equation and other rich calculation modes.
Power input: wide voltage design AC100V-240V 50-60Hz
Product quality: about 11Kg
Dimensions: 433mm(L)*320mm(W)*308mm(H)