Performance:
Sample capacity: 48 wells (6*8)
Reaction Volume: 15 - 100µL (15-50µL recommended)
Consumable: 0.2ml PCR tubes, 8-well strips, 48-well plates
Sensitivity: 1 copy
Linearity range: 1-10¹º
Reproducibility: CV<0.5% across 48 wells
Correlation coefficient: -0.99 to -1.00
Resolution: can discriminate between 1000 copies and 2000 copies
Dyes and probes: Ch1: FAM™/SYBR Green®, Ch2: VIC®/HEX™/JOE™/TET™, Ch3: CY5™
Hardware Thermal Unit:
Temperature control unit: semiconductor
Temperature control mode: tube control block control
Temperature range: 4~99°C
Max ramp rate: 4.0°C/s
Temperature accuracy: ±0.1°C
Temperature uniformity: ±0.1°C
Hot-lid: 30-108°C (default 105°C, adjustable) electronic automatic hot-lid
Optical Unit:
Light Source: LED (free of maintenance)
Detector: photosensor
Excitation: Ch1 470 nm, Ch2 530nm, Ch3 630nm, Ch4 custom-made, Ch5 custom-made, Ch6 custom-made
Emission: Ch1 510 nm, Ch2 565nm Ch3 665nm, Ch4 custom-made, Ch5 custom-made, Ch6 custom-made
Custom-made channels: Up to 6 channels, support 4-plex detection
Software:
Application: Absolute Quantification (AQ), Positive/Negative, Allelic Discrimination (TaqMan® endpoint method), Multiplex Melting Curve analysis, High Resolution Melting Curve analysis
Data export: Excel, CSV
Work condition:
Operation system: Microsoft ® Windows XP/Vista/7/8
Computer requirement: 1GB RAM, 40GB HD, CPU 1.5GHz
Power supply: 220V, 50HZ
Power consumption: 850VA
Communication: RS232 USB
Dimensions: 380mm×520mm×250mm (W×D×H)
Weight: 18kg