Analytical pump and fluidics:
Type: Serial dual-reciprocating pistons, microprocessor-controlled constant stroke, variable speed
Construction: Chemically inert, metal-free PEEK pump heads and flow paths compatible with aqueous eluents of pH 0–14 and reversed-phase solvents
Pump operating pressure: 0–35 MPa (0–5000 psi)
Flow rate range: 0.00–5.00 mL/min without changing pump heads
Flow precision: <0.1%, typically
Flow accuracy: <0.1%, typically
Pressure ripple: <1% at 13.8 MPa (2000 psi) and 1.0 mL/min
Eluent on-off valve: Standard
Piston seal wash: Dual-pump head, wash can be continuous when connected to rinse solution supply
Pressure alarm limits: Upper limit 0–35 MPa or 0–5000 psi in one unit (MPa or psi) increments; lower limit can be set up to one unit lower than upper limit
Vacuum degas: Yes, optional, automatic control
Eluent bottles: Polypropylene, up to 4L volume
Eluent bottle pressure: Not required
Columns supported: 2, 3, 4, and 5 mm i.d.; maximum length 250 mm analytical column with 50 mm guard column
Injection valve:
Type: 6-port, 2 position Rheodyne valve, electrically actuated
Column heater (optional):
Operating temperature range: 30 to 60 °C (86 to 140 °F); minimum 5 °C above ambient; settable range is equal to working range
Temperature accuracy: ±0.5 °C at sensor, at 40 °C
Suppressors and control:
Chemical suppression: 2 mm and 4 mm anion and cation, membrane suppression bed types
Displacement chemical suppression: 2 mm and 4 mm anion and cation, membrane suppression bed types
Electrolytic suppression: 2 mm and 4 mm anion and cation, membrane suppression bed types available
Electrolytic suppression with external water mode: 2 mm and 4 mm anion and cation, membrane suppression bed types available
Current control range: Thermo Scientific Dionex ERS 500 Electrolytically Regenerated Suppressor: 4 mm, 0–500 mA in 1 mA increments 2 mm, 0–150 mA in 1 mA increments
Thermo Scientific Dionex ERD 500 Electrolytically Regenerated Desalter: 0–500 mA (4 mm) and 0–150 mA (2 mm) in 1 mA increments
Thermo Scientific Dionex SRN Self-Regenerating Neutralizer: 0–500 mA in 1 mA increment
Salt converter: Available in 2 and 4 mm versions
Carbonic acid removal for anions: Dionex AERS 500 suppressor with Thermo Scientific Dionex CRD 200 Carbonate Removal Device for hydroxide eluents Dionex AERS 500 suppressor with Dionex CRD 300 device for carbonate eluents
Non-suppressed chromatography: Yes, supported
Suppressor wear parts: None; peristaltic pump and inline filters not required
Conductivity detector electronics and flow cell:
Type: Microprocessor-controlled digital signal processor
Cell drive: 8 kHz square wave
Detector linearity: 1% up to 1 mS
Maximum cell pressure: 10 MPa (1500 psi)
Resolution: 0.00238 nS/cm
Full-scale output ranges: Digital signal range 0–15000 μS/cm
Analog signal range 0–15000 μS/cm
Electronic noise: ±0.1 nS when background conductivity is 0–150 µS/cm
±2 nS when background conductivity is 151–3200 μS/cm
Filter: Rise times from 0 to 10 s, user selectable
Temperature compensation: Fixed at 1.7% per 1 °C at cell temperature
Temperature range: Ambient +7 °C, 30 to 55 °C
Cell electrodes: Passivated 316 stainless steel; compatible with MSA
Cell body: Chemically inert polymeric material
Cell volume: <1 µL
Heat exchanger: Inert, tortuous path for low axial dispersion
Autosampler:
Automation using autosampler: For more information about available autosamplers, visit the IC Autosamplers page
System software:
Chromeleon CDS, requires Microsoft® Windows® 10
Physical specifications:
Dimensions (h × w × d): 56.1 cm × 22.4 cm × 53.3 cm (22.1 in × 8.8 in × 21 in)
Weight: 24.5 kg (54 lb)
Power requirements: 100 to 240 V AC, 50 to 60 Hz, autoranging
Operating temperature: 4–40 °C (40–104 °F); cold-room-compatible (4 °C) as long as system power remains on
Operating humidity range: 5–95% relative, noncondensing
Control modes: Full control through Chromeleon CDS software; alternative control through TTL or relay closures; two relay outputs, two TTL outputs, four programmable inputs
USB communication protocol: One USB input; one built-in two-output USB hub
Leak detection: Built-in, optical sensor