Unlock Exclusive Discounts & Flash Sales! Click Here to Join the Deals on Every Wednesday!

Agilent 1260 Series - G1377A HPLC Well Plate Autosampler, Agilent Technologies (CAT#: STEM-C-0089-LGZ)

Highlights

1. The optimized design of the micro Rheodyne valve and needle seat, loop and seat capillaries minimizes dispersion.
2. High-resolution metering device provides ten times better resolution than standard autosamplers, convenient bypass operation Low delay volume, increased sample injection speed for high sample throughput, flexible and convenient sample handling of different types sample container.
3. Up to 768 samples can be processed unattended using a 384-well plate.

Cat Number: STEM-C-0089-LGZ

Application: For HPLC system.

Model: 1260 Series - G1377A

Add to Cart



Description

The manufacturer of Agilent 1260 Series - G1377A HPLC Well Plate Autosampler is Agilent Technologies. Model is 1260 Series - G1377A. The Agilent 1260 Infinity G1377A High Performance Micro Autosampler is designed to perform capillary LC with injection volumes from nL to μL.

Specification

Condition: used
GLP features: Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications: Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection range:
– 8 µL in 0.01 µL increments with the small loop capillary
– 40 µL in 0.01 µL increments with the extended loop capillary
Precision:
Typically < 0.5 % RSD of peak areas from 5 – 40 µL
Typically < 1 % RSD from 1 – 5 µL
Typically < 3 % RSD from 0.2 – 1 µL
Sample viscosity range: 2 – 5 cp
Sample capacity:
2× well-plates (MTP) + 10 × 2 ml vials
100 x 2 mL in one tray
40 x 2 mL in half tray
Injection cycle time:
Typically, < 30 s using following standard conditions:
Default draw speed: 4 µL/min
Default eject speed: 10 µL/min
Injection volume: 0.1 µL
Carry-over:
Typically < 0.05 % using the following conditions:
Column: 150 x 0.5 mm Hypersil ODS, 3 µm
Mobile phase: Water/Acetonitrile = 85/15
Column Flow rate: 13 µL/min
Injection volume: 1 µL caffeine (=25ng caffeine), 1 µL water to test carryover
Outside wash of needle before injection: 20 sec with water using flush port

Features

1. The optimized design of the micro Rheodyne valve and needle seat, loop and seat capillaries minimizes dispersion.
2. High-resolution metering device provides ten times better resolution than standard autosamplers, convenient bypass operation Low delay volume, increased sample injection speed for high sample throughput, flexible and convenient sample handling of different types sample container.
3. Up to 768 samples can be processed unattended using a 384-well plate.

Related Products

Agilent 1100 Series - G1367A HPLC Well Plate Autosampler, Agilent Technologies(CAT#-STEM-C-0052-LGZ)
Agilent 1100 Series - G1377A HPLC Well Plate Autosampler, Agilent Technologies(CAT#-STEM-C-0054-LGZ)
Agilent 1100 Series - G1389A HPLC Autosampler, Agilent Technologies(CAT#-STEM-C-0056-LGZ)
Agilent 1100 Series - G2260A HPLC Preparative Autosampler, Agilent Technologies(CAT#-STEM-C-0058-LGZ)
Agilent 1200 Series - G1329A HPLC Autosampler, Agilent Technologies(CAT#-STEM-C-0071-LGZ)
Agilent 1200 Series - G1367B HPLC Well Plate Autosampler, Agilent Technologies(CAT#-STEM-C-0074-LGZ)
Advertisement