3-way valve: Three-port, PTFE, 67 µL internal volume. 3,5 µL dead volume from the common port to the normally closed port, up to 20 mL/min with large bore needle OR Three-port, PTFE, 114 µL internal volume. 6 µL dead volume from the common port to the normally closed port, up to 200 mL/min with large bore needle.
Contact control
Contact closure inputs for start/advance and end/home
Event mark output
One programmable output for control of peripheral devices
Data acquisition: An input channel for analog-to-digital conversion; digital data is transmitted on the GSIOC
Detector input: ±100 mV full scale
Display: two lines of 24 alphanumeric characters; backlit LCD
Drop counting: Up to 9999 drops per fraction. Maximum rate: 20 drops per second
Environmental conditions
Indoor use only
Up to 2000 meters altitude
5° - 40 °C temperature range
75 - 105 kPa air pressure
Pollution degree 1 or 2 in accordance with IEC 66
Maximum relative humidity 80% for temperatures up to 31°, decreasing linearly to 50% relative humidity at 40 °C.
Event marker: 100 ms pulse (contact closure)
Maximum collection volume/tube: 32 mL (18 x 180 mm tube) or 25 mL (18 x 150 mm tube)
Maximum fractions: up to 768 fractions when configured with eight 96-well, 2 mL Whatman/Plyfiltronics microplates on a Code 17 rack
Multi-cycle operation: repetitive collection of each sample into same set of tubes or collection each sample into different set of tubes
Multiple column collection: simultaneous collection from up to 18 columns with installation of optimal multiple column adapter(s).
Number of racks: One Code 16 or 16D rack, one Code 17 rack, or up to four Code 20-Series racks (all racks must be same code)
Operating modes: Time, drop, peak + time, peak + drop, and manual
Peak detection
Adaptive slope algorithm that applies use-specified parameters to accommodate drifting baselines, negative peaks, and asymmetrical peaks or absolute threshold level that collects all peaks above specified millivolt value
Peak Parameters: peak height or level, in mV, and peak width at half-height, in minutes
Minimum peak height for fractionation: 0.1 mV full scale
Detector-collector delay: 0.01 minute increments
Physical space requirement: 47.9 x 46.4 x 33.0 cm (18.9 x 18.3 x 13.0 in.)
Power requirements
Frequency: 50 to 60 Hz
Voltage: 100 - 120 V or 220 - 240 V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 0.5 A for 100 - 120 V or 0.25 A for 220 - 240 V
Programmable time units (min.): from 0.01 to 99.99 minutes per tube, with 0.01 minute limit of resolution
Software control: Via Gilson Serial Input/Output CHannel (GSIOC) or by contact closure
Time based programming: Up to ten collection windows and ten drain steps in any mode
Tube change time: 100-720 ms. Center-to-center, depending on rack type
Weight: 10.9 kg (24 lbs)