The manufacturer of the HP 8452A Diode Array UV-Vis Spectrophotometer is Agilent Technologies/HP. Refurbished Hewlett Packard(HP) model 8452A diode array UV-Vis spectrophotometer complete with cell holder and HP G1115AA General Purpose software Rev A.02.04.
Specification
Condition: Refurbished, Certified Pre-owned Wavelength Range: 190 to 820 nm Wavelength Accuracy: < ± 2 nm Wavelength Reproducibility: ± 0.05 nm (typical with constant conditions) Spectral Bandwidth: 2 nm Stray Light: at 200 nm = < 0.05% , at 340 nm = < 0.05% Photometric Range: 0.0022 – 3.3 Å; 0.5 second measurements at 350 nm Photometric Accuracy: ±0.005 Å; at 1 Å, 440 nm, NIST 930e Photometric Noise: <0.0002 Å RMS; at 0 Å, 500 nm, sixty 0.5-second scans Photometric Stability: <0.001 Å/hr; (at 0 Å, constant ambient conditions, after 1 hour warm up): measured over 60 second period, every 5 seconds Baseline Flatness: <0.001 Å RMS; 0.5 second blank followed by 0.5 second scan Shortest Scan Time: 0.1 seconds; full range Rapid Scan Mode: 10 scans per second; 70 nm range Power Requirements: Line Voltage: 80-132 or 170-264 V; AC Line Frequency: 47-66Hz Dimensions (L x W x H): 23.8 in x 16.7 in x 7.3 in Weight: 33.2 lbs.
Features
Refurbished to meet manufacturer's specification In excellent condition Thoroughly cleaned Money back guarantee Certified with warranty