Thermo Lab Systems Multiskan Ascent 354 Microplate Photometer, Thermo Fisher Scientific (CAT#: STEM-SPO-0384-ZJF)

Highlights

Excellent optical performance for 96- and 384-well plates:
The high-quality optical system of the Multiskan Ascent ensures excellent reproducibility, accuracy and reliable results.
Fast measurements:

Cat Number: STEM-SPO-0384-ZJF

Application: To measure chemical, biological or physical reactions, properties, and analytes within the well of a microplate




Description

The manufacturer of the Thermo Lab Systems Multiskan Ascent 354 Microplate Photometer is Thermo Fisher Scientific.
The Thermo Labsystems Multiskan Ascent is an advanced, compact, one-channel vertical light path filter photometer designed to perform standard photometric measurements. The measurement types include end point, two point and kinetic measurements.

Specification

Condition: Pre-owned
Optical performance, wavelength range: 340-850 nm
Filters: 8-position filter wheel with 405, 450, 490 and 630nm or filters of choice
Half bandwidth of filters: 3-9 nm
Wavelength accuracy: +/- 2 nm
Light source: Tungsten halogen lamp
Detector: One silicon photo detector
Linearity (96-well plate):
0-4 Abs, +/- 2 %
0-3 Abs, +/- 2 %, 340 nm and 405 nm
0-3 Abs, +/- 2 % with continuous mode
Linearity (384-well plate):
0-3 Abs, +/- 2 %
0-2 Abs, +/- 2 % with continuous mode
Read-out range: 0-6 Abs
Resolution: 0.001 Abs
Accuracy:
+/- 1 % or 0.003 Abs, whichever is greater (0-2 Abs)
+/- 2 % (2-3 Abs)
Precision:
CV < 0.2 % (0-3 Abs),
CV < 1.0 % (3-4 Abs) stepping mode
CV < 0.3 % (0-2 Abs),
CV < 1.2 % (2-3 Abs) continuous mode
Measurement time (96-well plate):
Continuous mode: 9 s
Stepping mode: 14 s
Measurement time (384-well plate):
Continuous mode: 20 s
Stepping mode: 42 s

Features

Excellent optical performance for 96- and 384-well plates:
The high-quality optical system of the Multiskan Ascent ensures excellent reproducibility, accuracy and reliable results.
Fast measurements:
Multiskan Ascent provides not only extreme accuracy and precision, but also very rapid measurements: It takes only 9 seconds to measure a 96-well plate and only 20 seconds to measure a 384-well plate. These values can have a noticeable effect on a laboratory’s throughput.
Easy and flexible robot integration:
Another feature that makes Multiskan Ascent ideal for high-throughput environments is its simple and easy robot integration. This unit is designed for integration, which can increase walk-away reading capacity to 20 or 80 microplates.

Related Products

Thermo Scientific FH100M Peristaltic Pump, Thermo Fisher Scientific
Digital 2 Heat Block 120, VWR
GeneMate BT1502 Orbital Shaker, VWR
Advertisement