ZNCL-BS Intelligent Digital Display Magnetic Heating Plate, 100-2500rpm, model ZNCL-BS is manufactured by Gongyi Yuhua Instrument Co., Ltd. The method of use of this product is:
1. Fix the vertical rod in the screw hole on the back of the stirrer, adjust the height of the cross clip, fix the reaction bottle with the universal clip, put in a suitable stirrer, plug in the internal sensor plug or external sensor probe, and plug in the power supply ~220V, turn on the main switch.
2. Temperature setting: Press the heating switch button to enter the heating state, and the PV and SV windows have digital display. Press the SET button, the numbers in the SV window flash, and the numbers in the SV window can be adjusted by pressing the "<", "∨", and "∧" buttons. Press the SET key again to exit the SV window temperature setting state, and the SV window temperature setting is completed. Set the required heating temperature, such as: 100 ℃, the green light is on to indicate heating, and the green light is off to indicate stop. The microcomputer will determine the heating amount according to the temperature difference between the set temperature and the current temperature, to ensure that there is no temperature rush to heat up in place at one time, and to maintain the balance of supply and heat dissipation under the temperature difference between the set value and the displayed value ±1 °C, so that the heating process can be completed easily.
3. Speed setting: Press the stirring switch button to enter the stirring state, and the r/min window has a digital display. Press the SET button, the numbers in the SV window flash, and then press the SET button again, the SV window setting status exits, and the numbers in the r/min window flash. ” button to adjust. Press the SET button again to exit the r/min window speed setting state, and the r/min window speed setting is complete. 100-2500 rpm, LCD display, key selection, stable at low speed, strong at high speed. +
4. Self-tuning function: When the self-tuning function is activated, the heating time is the shortest, the flushing temperature is the smallest, and the balance is the best when different heating sections or heating power and solution are irregular. However, the self-tuning should be restarted after changing the heating medium or heating conditions set up.
5. Start auto-tuning: in the normal measurement control state, press and hold the "<" key for 8 seconds to enter the auto-tuning state. The AT light flashes in the auto-tuning state. If the process needs to stop the auto-tuning, press and hold the "<" key again. 8 seconds is enough.
6. When pressing the SET key to adjust each function parameter, if there is no key operation within 8 seconds, the instrument will automatically exit the setting state and enter the normal display state. If the parameters still need to be adjusted, you need to press the SET key again to enter the setting state.
7. There is a rubber plug under the back of the stirrer to protect the external thermocouple socket from corrosion and rust and to conduct the inner wire. If it is pulled out, the inner probe will be disconnected and the machine will stop working. If the external thermocouple is used, the plug should be removed for storage, the external thermocouple plug should be inserted into the socket and the nut should be tightened, and then the stainless steel probe should be placed in the solution for temperature control heating.
8. The electric appliance has the function of breaking couple protection. When the thermocouple is poorly connected, the green light of the display window "hhhh" will go out, and the electric appliance will stop heating, and it needs to be checked before use.