The manufacturer of the Agilent 1200 G1314B VWD Variable Wavelength Detector, Agilent Technologies is Agilent Technologies.
Specification
Condition: Refurbished, Certified Pre-owned Detection type: Double-beam photometer Light source: Deuterium lamp Noise: ± 0.75 x 10 -5 AU, at 254 nm Drift: 3 x 10-4AU/hr, at 254 nm Linearity: >2 AU upper limit Wavelength range: 190-600 nm Wavelength accuracy: ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter Wavelength bandwidth: 6.5 nm typical over whole wavelength range Maximum data rate: 13 Hz Flow cells: Standard: 14-µL volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum High Pressure: 14-µL volume, 10-mm cell path length and 400 bar (5880 psi) Pressure maximum Micro: 1-µL volume, 5-mm cell path length and 40 bar (588 psi) pressure maximum Semimicro: 5-µL volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum Communications: Controller-area network (CAN), RS-232. APG Remote: ready, start, stop and shut-down signals, LAN optional Safety and maintenance: Extensive diagnostics, error detection and display through control module and Agilent ChemStation, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. Glp: Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter. Housing: All materials recyclable
Features
Refurbished to meet manufacturer's specification In excellent condition Thoroughly cleaned Money back guarantee Certified with warranty